Terminated

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

FDA Recall Z-0847-2015 · Posted 2014-12-22

Recall StatusTerminated
Product CodeLBI
Date Initiated2014-11-19
Date Terminated2015-07-20
Product Quantity1907 total (1413 in US and 494 international)

Reason for Recall

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Firm Action

Natus Medical has begun mailing the Urgent Recall Notification/Field Safety Notice neoBLUE LED Board Replacements letter (DOC-012737B), dated November 5, 2014, to US consignees. Natus Medical also send the Product Advisory neoBLUE LED Board Replacements letter (DOC-0012738), dated October 30, 2014, to International Distribution Partners. Both US and International consignees also received the TECHNICAL BULLETIN LED Board Kit, neoBLUE 2 LED Phototherapy System, p/n: 001840, Treatment Intensity with revised LED board for neoBLUE 2 systems (p/n: 040904 & 040906) [DOC 008353]; and the Label, neoBLUE Intensity/Height (DOC 011561). Consignees are instructed to measure the light intensity of their devices and to document that information of the device using an adhesive label that is provided. The neoBLUE 3 system and its replacement panel sold today are not affected by this issue. If there are any questions regarding this advisory letter or the attached Technical Bulletin, please call Natus Technical Service at 888-496-2887.

Root Cause

Other

Distribution Pattern

worldwide