Terminated

Video Cart, Premium 120V product labeling is as follows: REF, VP6501, QTY1, Video Cart, Premium, 120V, Rx ONLY. These Video Carts are intended for storage of Endoscopic equipment that is used in single or multiple surgical specialties MANUFACTURED FOR: Linvatec, LINVATEC CORPORATION, LARGO, FL 33773 USA

FDA Recall Z-0848-2011 · Posted 2011-01-04

Recall StatusTerminated
Product CodeKQM
Date Initiated2010-02-23
Date Terminated2012-03-22
Product Quantity333 carts

Reason for Recall

ConMed Linvatec, Endoscopy Division, dba. ConMed Linvatec, Largo, Florida is recalling five products: (1.) Video Cart Premium 120 VAC (VP6501), (2.) Video Cart Multi-Specialty 120 VAC (VP6502), (3.) Video Cart Articulating Arm 120V (VP6504), (4.) Video Cart (VP8500), and (5.) Power Cord Assembly Self Retracting (VP6547). These products are being recalled due to there is a possibility that the Vid

Firm Action

ConMed Linvatec sent URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters to their customers by FedEx. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to review their plugs for damage and determine if they were affected. Distributors were asked to contact affected customers and schedule delivery of their replacement plugs within 30 days. If the account no longer has the product in their possession, or if they have transferred the product, they were asked to respond as such via a Reply form. Customers were asked to document delivery of replacement plugs on the Sales Rep Reply Form and mail back or return by fax to 1-888-370-3062. For questions regarding this recall call the Customer Service Department at 1-888-292-0100 or e-mail [email protected].

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, AND WY and the countries of Canada, Chile, Colombia, Denmark, Korea, Mexico, Singapore, South Africa, United Arab Emirates, and Venezuela.