Prima III TotalSENSE Torso Coil (Body Array Coil), labeled as Manufactured by: USA Instruments, Inc. This device is a component of the Magnetom TRIO 3.0T MRI Systems, manufactured by Siemens AG Medical Solutions.

FDA Recall Z-0849-04 · Posted 2004-07-20

Recalling FirmUSA Instruments Inc.
Recall StatusTerminated
Product CodeLNH
Date Initiated2004-03-19
Date Terminated2012-03-01
Product Quantity79

Reason for Recall

A defective capacitor in the Body Array Coil may result in unstable, unreliable performance, with excessive heating of patient's body.

Firm Action

The firm notified the regional Siemens Service Centers by letter on 3/19/2004. The affected coils will be returned to the recalling firm for modification.

Root Cause

Other

Distribution Pattern

The device was shipped to domestic consignees located CT, NY, PA, NC, GA, FL, OH, IN, MN, IL, UT, CA, OR; and foreign consignees in Austria, Denmark, Spain, Hong Kong, China, Netherlands, Japan, Taiwan, Korea, Australia, Turkey, Brazil, Germany.