Cybernet Medical MedStar Telemedicine System, Manufactured by Cybernet Medical, Ann Arbor, MI; Model 533-210. The device is for a remote medical data collection device for the monitoring of patients by healthcare providers. It is intended to work with a weight scale, spirometer, blood pressure unit and pulse oximeter and to transmit data over a phone line.

FDA Recall Z-0852-2010 · Posted 2010-02-25

Recall StatusTerminated
Product CodeDXH
Date Initiated2008-01-23
Date Terminated2010-10-22
Product Quantity60

Reason for Recall

A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.

Firm Action

The firm sent out a recall letter dated 10/27/09. Consignees were asked to return the devices in prepaid shipping containers. The recalling firm would do a software upgrade and return the device to the consignee. The firm expects it to take 3-6 months to upgrade all devices. Contact Cybernet Medically immediately, at 1-800-CYBERNET, with the serial number and code versions of the devices in question and Cybernet will start the RMA process for returning the devices for the updates, free of cost.

Root Cause

Component design/selection

Distribution Pattern

Nationwide distribution: California, Florida, Georgia, Illinois, Louisiana, Michigan, Mississippi, New York, Oklahoma and West Virginia.