GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG
FDA Recall Z-0853-04 · Posted 2004-07-20
Advertisement
Recalling FirmGeneral Electric Med Systems
Recall StatusTerminated
Product CodeEAM
Date Initiated2004-03-30
Date Terminated2006-07-24
Product Quantity1236 devices
Reason for Recall
The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Firm Action
Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
Root Cause
Other
Distribution Pattern
The units were distributed nationwide in the United States and worldwide.