Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG

FDA Recall Z-0853-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeEAM
Date Initiated2004-03-30
Date Terminated2006-07-24
Product Quantity1236 devices

Reason for Recall

The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.

Firm Action

Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.

Root Cause

Other

Distribution Pattern

The units were distributed nationwide in the United States and worldwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑