STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.
FDA Recall Z-0853-2007 · Posted 2007-07-03
Advertisement
Recalling FirmDiagnostica Stago, Inc.
Recall StatusTerminated
Product CodeGGP
Date Initiated2007-03-13
Date Terminated2009-01-16
Product Quantity1,098 units
Reason for Recall
A dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.
Firm Action
Customers were notified on March 13, 2007 by letter, first class mail.
Root Cause
Other
Distribution Pattern
Hospitals and laboratories Nationwide