Terminated

Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

FDA Recall Z-0854-2015 · Posted 2015-01-14

Recall StatusTerminated
Product CodeIZI
Date Initiated2014-09-17
Date Terminated2015-08-06
Product Quantity44

Reason for Recall

The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.

Firm Action

Planned action (s) to repair defect or to bring product into compliance for Siemens Medical Solutions Inc. concerning the corrective action plan (CAP) for the Model(s) Artis Zee/Zeego Stationary Angiographic X-ray Systems. 1.Notifying the customer of the defect. 2.Implementing a corrective software patch to protect the system from damage in the event of hardware failure 3. Scheduling a service visit from a Siemens representative to perform a hardware replacement. 4. Replacing the non-conforming X-ray tube with one manufactured after the nonconformity issue was resolved. For further questions please call (610) 219-6300.

Root Cause

Process control

Distribution Pattern

USA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.