UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access Clinical Systems with Dxl System Software versions 4.2 or 4.2.1. In vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentrations found in human body fluids.
FDA Recall Z-0856-2010 · Posted 2010-02-25
Advertisement
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2009-10-28
Date Terminated2012-05-07
Product Quantity377 Instruments
Reason for Recall
This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample processing, the system will post a yellow caution event and continue to operate with the dispense plate in an elevated position until the instrument
Firm Action
Beckman Coulter issued an "Urgent: Product Corrective Action" letter dated October 28, 2009. Customers were notified of the affected product and how the issue will be resolved.
For further information, contact Beckman Coulter Inc. Technical Support at 1-800-854-3633.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution -- United States and Canada.