Terminated

RTH8 Rotor, P/N X01-005847-001. RTH8 used in the StatSpin Express 4 Horizontal Centrifuge, Model M510, Product No. SSH4. The RTH8 rotor is used with the Statspin Express 4 Horizontal Centrifuge. StatSpin Express 4 Centrifuge: For in vitro diagnostic use to produce the rapid separation of whole blood contained in original collection tubes.

FDA Recall Z-0857-2015 · Posted 2014-12-23

Recalling FirmIris Diagnostics
Recall StatusTerminated
Product CodeJQC
Date Initiated2014-12-03
Date Terminated2015-11-02
Product Quantity3912 units

Reason for Recall

Iris International is recalling the RTH8 Rotor used in the StatSpin Express 4 Horizontal Centrifuge because the RTH8 rotor may develop cracks with use over time.

Firm Action

A customer notification letter dated 12/3/14 was sent to all customers to inform them that Iris International is recalling the RTH8 Rotor used in the StatSpin Express 4 Horizontal Centrifuge because the RTH8 rotor may develop cracks with use over time. The letter informs the customers of the problems identified and the actions to be taken. Beckman Coulter will be managing the logistics of the recall notice. Customers with questions are instructed to contact Customer Technical Support at (800) 854-3633 or via website at http://www.beckmancoulter.com.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Norway, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Venezuela.