Cyberonics VNS Therapy AspireHC, Model 105 Generator. The Cyberonics VNS Therapy System consists of the implantable VNS Therapy Generator, the VNS Therapy Lead, and the external programming system used to change the stimulation settings. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.

FDA Recall Z-0858-2015 · Posted 2014-12-23

Recalling FirmCyberonics, Inc
Recall StatusTerminated
Product CodeLYJ
Date Initiated2014-11-20
Date Terminated2016-02-08
Product Quantity1 unit

Reason for Recall

The recalled product was distributed with an incorrect serial number printed on the device's label.

Firm Action

The recalling firm visited the one consignee on 11/4/2014 to inform the consignee of the recall and arrange for the return of the recalled devices.

Root Cause

Labeling mix-ups

Distribution Pattern

Distributed to PA.