Infant Nasal CPAP Cannula, Size 2; an Rx infant nasal prong continuous positive airway pressure (CPAP) cannula one inspiratory and one expiratory elbow connector; 10 units per case; Teleflex Medical, Research Triangle Park, NC 27709 U.S.A., Made in Mexico; The cannulas are sold as components in the following configurations: a) catalog No. 1692: Infant CPAP Set intended for use with mechanical ventilators, which includes one cannula with one inspiratory and one expiratory elbow connector, two 10mm to 7.5mm adapters, one knit cap and two 6 inch sections of Velcro securing tape; b) catalog No. 1686: Infant CPAP System intended for use with a water seal or spring loaded valve, and includes one cannula with one inspiratory and one expiratory elbow conenctor, two 4 ft. lengths of 10mm I.D. corrugated tubing, one 4 ft. pressure monitoring line, one 22mm to 10mm humidifier adapter, one knit cap and two 6 inch sections of Velcro securing tape;

FDA Recall Z-0859-2007 · Posted 2007-06-07

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBZD
Date Initiated2007-03-23
Date Terminated2009-11-16
Product Quantity13,530 for all products

Reason for Recall

The luer-port in the expiratory elbow may be occluded, which interferes with a ventilator's ability to detect an accurate pressure reading.

Firm Action

Teleflex Medical sent Urgent Medical Device Recall letters dated 3/23/07 via UPS 2nd Day Air to the direct accounts, informing them that the luer-port in the expiratory elbow may be occluded, which interferes with a ventilator's ability to detect an accurate pressure reading. The accounts were requested to cease use of the affected units, remove them from use/stock, quarantine and destroy them. If the accounts further distributed the product, they were requested to conduct a sub-recall of the affected units from their customers. The accounts were instructed to complete and return by fax the enclosed Acknowledgement & Stock Status Form, indicating the number of units destroyed and if they are requesting replacement product. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Puerto Rico, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, El Salvador, Germany, Guatemala, Hong Kong, India, Israel, Kuwait, Malaysia, Mexico, New Zealand, Paraguay, Peru, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, and Venezuela

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →