Terminated

Specify 5-6-5 Lead, Specify 2x8 Lead, Model Numbers: 39565, 39286. Spinal cord stimulation leads. Part of a neurostimulation system for treating chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.

FDA Recall Z-0859-2016 · Posted None

Recall StatusTerminated
Product CodeLGW
Date Initiated2016-01-11
Date Terminated2017-01-20
Product Quantity51,138 (44,006 US; 9,132 OUS)

Reason for Recall

Some Specify 5-6-5 and 2x8 surgical lead product labeling does not adequately distinguish between permanent surgical implant and trialing/screening use.

Firm Action

Consignees were sent a Medtronic "Medical Device Correction" letter dated January 2016. The letter was addressed to Healthcare Professional. The letter described the problem and the product involved in the recall. Advised consignees to ensure their staff of the action that is being recommended. Requested consignees to complete and return the "Physician Confirmation Form". For additional information they can contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekends 7 am - 6 Pm Central time.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution, including US Nationwide.

Official Source

View FDA Recall →