Open, Classified

SIGNA Excite 3T. Magnetic Resonance Imaging System

FDA Recall Z-0859-2022 · Posted 2022-04-04

Recalling FirmGE Healthcare, LLC
Recall StatusOpen, Classified
Product CodeLNH
Date Initiated2022-02-18
Product Quantity55 units

Reason for Recall

There is potential for the images to be flipped left to right.

Firm Action

This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑