Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system.
FDA Recall Z-0860-2012 · Posted 2012-01-25
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeMQB
Date Initiated2011-12-12
Date Terminated2014-01-14
Product Quantity1
Reason for Recall
Fixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type.
Firm Action
Siemens Healthcare sent a "SAFETY ADVISORY NOTICE" dated December 12, 2011 to the affected customer. The letter identifies the product, problem, and actions to be taken by the customer. The firm prepared a modification, which will be provided to the affected customer under Update Instruction XP055/11/S. The firm will resolve this potential issue by preventively exchanging all bolts for the fixation of the X-ray tube holder of the concerned AXIOM Aristos FX Plus unit.
Root Cause
Process control
Distribution Pattern
Product distributed in one hospital in California.
Related FDA 510(k) Records