Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.

FDA Recall Z-0860-2021 · Posted None

Recall StatusOpen, Classified
Product CodeLGW
Date Initiated2020-11-12
Product Quantity7960 downloads

Reason for Recall

A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.

Firm Action

Medtronic notified the first consignee by letter on 11/12/2020 with a confirmation of their complaint and request to perform a correction to the device which put the device in a factory-reset state and restore functionality to allow programming and continued therapy. On 17 Dec 2020, Medtronic will begin issuing an Urgent Medical Device Correction notification to physicians regarding this validation error and providing instructions on the impact and appropriate mitigating actions if this issue presents.

Root Cause

Software design (manufacturing process)

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Canada, Latin America, Asia Pacific, Japan, China, and European regions.