Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.
FDA Recall Z-0864-05 · Posted 2005-06-02
Advertisement
Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2005-02-09
Date Terminated2005-06-29
Reason for Recall
The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
Firm Action
The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
Root Cause
Other
Distribution Pattern
Nationwide and Canada.