Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.

FDA Recall Z-0864-05 · Posted 2005-06-02

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2005-02-09
Date Terminated2005-06-29

Reason for Recall

The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.

Firm Action

The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.