Terminated

i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer

FDA Recall Z-0865-2010 · Posted 2010-02-26

Recall StatusTerminated
Product CodeJQP
Date Initiated2009-09-29
Date Terminated2010-07-01
Product Quantity36

Reason for Recall

Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of comment being included in the CDS test record only.

Firm Action

Urgent Recall Notification letters were sent by Fed Ex to all US customers on September 29, 2009 and worldwide on October 1, 2009. Questions should be directed to Abbott Point of Care Technical Support at 1-800-366-8010, Option 1.

Root Cause

Software design

Distribution Pattern

Nationwide and Foreign: Germany, the United Kingdom, Switzerland and the Netherlands.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →