VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.
FDA Recall Z-0866-05 · Posted 2005-06-04
Advertisement
Recalling FirmOrtho-Clinical Diagnostics
Recall StatusTerminated
Product CodeLEG
Date Initiated2005-04-18
Date Terminated2005-10-13
Product Quantity139 boxes
Reason for Recall
Negatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.
Firm Action
Letters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.
Root Cause
Other
Distribution Pattern
37 U.S. consignees nationwide. Ortho-Clinical foreign affiliates in England, France and Spain.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.