EXXCEL Soft Standard Wall EPTFE Vascular Graft; D:8 mm: L: 50 mm straight Maquet Cardiovascular LLS, 45 Barbour Pond Drive, Wayne, NJ 07470
FDA Recall Z-0866-2010 · Posted 2010-02-26
Advertisement
Recalling FirmMaquet Cardiovascular, LLC
Recall StatusTerminated
Product CodeDSY
Date Initiated2010-01-20
Date Terminated2010-07-06
Product Quantity18 units
Reason for Recall
Labeling mix up: Shelf Carton label for Exxcel Soft Thin Wall Vascular Grafts, lot # 12633379 were mislabeled as Exxcel Soft Standard Wall Grafts. Bleeding could occur if wrong product is used.
Firm Action
Recall letters were sent to all US customers on January 20, 2010 by certified mail. Letters for foreign consignees were to be sent as soon as translation was completed. Customer are directed to call the company at 973-709-7515 for more information.
Root Cause
Labeling mix-ups
Distribution Pattern
Product was distributed to hospitals in TX, TN, CT, AR, AZ as well as to Austria and France.