Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.

FDA Recall Z-0867-2021 · Posted None

Recall StatusTerminated
Product CodeGXP
Date Initiated2019-06-06
Date Terminated2024-09-18
Product Quantity49 units

Reason for Recall

A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.

Firm Action

On June 7, 2019, customers were notified by letter sent by FedEx that Skeletal Kinetics LLC (Skeletal Kinetics) is voluntarily recalling certain OsteoVation, Inject Bone Void Filler kits distributed by OsteoMed LLC.The recall was initiated to correct a discrepancy between the Mixing Instruction Card and the Instructions for Use. The letter notes that Skeletal Kinetics has not received any customer complaints or reports of injury to the patient or end user. Customers are requested to: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or the enclosed pre-paid envelope; and (2) return unused inventory via Federal Express with reference to Return Material Authorization FER 3003890476-6/6/19-001-R or by pickup by the sales representative indicated on the Acknowledgement and Receipt Form. Customers with questions are instructed to contact Mr. Randy Mosley, Director of QA, Skeletal Kinetics LLC by telephone at 972 677 4789 Monday-Friday, 8 AM 5 PM CDT.

Root Cause

Labeling Change Control

Distribution Pattern

U.S. Nationwide distribution including in the states of TX, NY, FL and AZ.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑