Roche/Hitachi Modular Analytics System, Modular D Module GMMI clinical chemistry analyzer; Catalog number 03284522001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests. Roche Diagnostics Corporation, 9115 Hague Road, Indianapolis, IN, 46250

FDA Recall Z-0869-2008 · Posted 2008-08-13

Recall StatusTerminated
Product CodeJJE
Date Initiated2008-03-21
Date Terminated2009-11-18
Product Quantity11 of all models

Reason for Recall

Possible mismatch between patient and result.

Firm Action

Consignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.

Root Cause

Process change control

Distribution Pattern

Nationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.