Elecsys Troponin T; catalog number 2017644

FDA Recall Z-0870-03 · Posted 2003-05-29

Recall StatusTerminated
Product CodeMMI
Date Initiated2003-04-25
Date Terminated2003-06-19

Reason for Recall

Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.

Firm Action

An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.

Root Cause

Other

Distribution Pattern

United States