ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ.
FDA Recall Z-0870-06 · Posted 2006-05-17
Advertisement
Recalling FirmExcelsior Medical Corp
Recall StatusTerminated
Product CodeFRN
Date Initiated2006-03-21
Date Terminated2006-12-07
Product Quantity3911 units
Reason for Recall
Potential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a 140cc syringe. The ESP Pump works correctly when used in 60 mode with a 60cc syringe or smaller.
Firm Action
An 'Urgent Recall' letter was mailed certified return receipt mail on 3/22/06 to all customers per distribution list.
Root Cause
Other
Distribution Pattern
Nationwide and Quebec, Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.