Elecsys Troponin T STAT; catalog number 2017423
FDA Recall Z-0871-03 · Posted 2003-05-29
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeMMI
Date Initiated2003-04-25
Date Terminated2003-06-19
Reason for Recall
Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.
Firm Action
An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.
Root Cause
Other
Distribution Pattern
United States
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement