PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN.Catalog # 1711. Intended to provide a quantitative measurement of total cholesterol in whole blood.
FDA Recall Z-0871-2010 · Posted 2010-03-01
Advertisement
Recalling FirmPolymer Technology Systems, Inc.
Recall StatusTerminated
Product CodeCHH
Date Initiated2009-04-02
Date Terminated2010-10-20
Product Quantity486 vials
Reason for Recall
The memory chip was coded with incorrect product information.
Firm Action
Polymer Technology Systems, Inc. issued an "Urgent - Field Correction" letter dated April 2, 2009. Consignees were advised of the problem and asked to identify how many kits were in inventory so that the firm could replace product.
For further information, contact Polymer Technology Systems, Inc. Customer Service at 1-877-870-5610.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution