Zimmer Cannulated Screw System Cannulated Driver, 5.0 mm Hex, REF 1147-49-50; Catalog # 00-1147-049-50. Manufactured by Pioneer Surgical Technology, Marquette, MI; Distributed by Zimmer, Warsaw, IN.

FDA Recall Z-0872-06 · Posted 2006-05-18

Recall StatusTerminated
Product CodeHWC
Date Initiated2006-03-03
Date Terminated2007-12-04
Product Quantity74

Reason for Recall

There may be an obstruction in the cannula that will prevent the guide wire from passing through it. The obstruction in the cannula can force the guidewire to advance further into the patient than intended.

Firm Action

Pioneer notified Zimmer via letter dated 3/3/06 and Zimmer notified their hospital customers and distribution sites via letters dated 3/10/06 to check their inventory and return the product immediately to Zimmer.

Root Cause

Other

Distribution Pattern

Nationwide.