NeuViz 64 Multi-slice CT Scanner Systems

FDA Recall Z-0875-2017 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-08-31
Date Terminated2018-01-04
Product Quantity23 units

Reason for Recall

The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.

Firm Action

Neusoft Medical Systems Co. Ltd. planned action to bring into compliance. 1.Affected Customers will be notified by letter to provide them a description of the affected device, instructions for using the device prior to correction, hazards associated with the defect, and a statement of corrective actions to be taken at no cost to the owner. 2.Field Service Engineers will visit owner sites to install a software update which resolves the defect. 3.Corrections will be made at no cost to the owner 4.Corrections will be completed by May 11, 2017. CDRH approves the CAP subject to the following conditions: .Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. Further questions please call (281) 453-1205.

Root Cause

Software design

Distribution Pattern

US Distribution