Reason for Recall
Complaints about batteries indicating a satisfying remaining time but after a short time suddenly shut off without giving alarms to indicate short remaining battery back-up time. (Power off alarm will be given).
Firm Action
Maquet sent out a Safety alert Letter on 3/31/2005 to all users of the Servo-i.
Root Cause
Other
Distribution Pattern
The ventilators were distributed nationwide to hosptials and clinics.