LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

FDA Recall Z-0878-2020 · Posted None

Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeMBH
Date Initiated2019-12-16
Product Quantity59 units

Reason for Recall

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Firm Action

The firm initiated the recall by letter on December 16, 2019. The letter identified the problem and directed the consignee to remove the affected lots noted in the Urgent Medical Device Recall Notice.

Root Cause

Process change control

Distribution Pattern

USA: IA, NC, PA, TN. Int'l: CH, BE, GB, AE, CN, DE, NL, BR, FR.