GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments
FDA Recall Z-0880-2017 · Posted None
Advertisement
Recalling FirmStryker Howmedica Osteonics Corp.
Recall StatusTerminated
Product CodeNDE
Date Initiated2016-11-09
Date Terminated2019-03-27
Product Quantity2,084 units
Reason for Recall
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.
Firm Action
Stryker sent an Urgent Medical Device Removal letter dated November 9, 2016, to all affected customers via UPS (with return receipt). Stericycle will be handling the returns. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inform users of the Medical Device Recall and forward the notice to all those individuals who need to be aware within their organization. Return all affected products available at their location to:
Stryker C/O Stericycle
2670 Executive Dr., Suite A
Indianapolis, IN 46241
Attn: RA 2016-169 - Event XXX
Customers were also instructed to complete and sign the enclosed Business Reply Form and return to Stericycle.
Customers with questions should call +1-201-851-6683.
For questions regarding this recall call 201-831-5000.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution - US (nationwide) Internationally to AU, CA, CL, CH, ES, FR, GB, SE, JP, FR, ZA, and NL.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement