Roche brand AccuData GTS System; Catalog number 404, material number 03136833001.
FDA Recall Z-0882-05 · Posted 2005-06-11
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeCGA
Date Initiated2005-04-18
Date Terminated2006-03-07
Reason for Recall
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Firm Action
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Root Cause
Other
Distribution Pattern
Nationwide and Guam.