A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.
FDA Recall Z-0882-2021 · Posted 2021-01-21
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeHRS
Date Initiated2020-12-23
Date Terminated2023-04-13
Product Quantity174 systems
Reason for Recall
The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
Firm Action
On 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
Root Cause
Device Design
Distribution Pattern
Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.