Discovery Multiprogrammable Pacemaker, DDDR Model 1273
FDA Recall Z-0883-03 · Posted 2003-05-29
Advertisement
Recalling FirmGuidant Corp-Cpi Division
Recall StatusTerminated
Product CodeDXY
Date Initiated2003-05-06
Date Terminated2005-11-18
Reason for Recall
In a rare circumstance (fallback mode when high-rate atrial activity is detected), the battery gauge can over-estimate battery life.
Firm Action
An 'Important Pacemaker Follow-up Information' letter dated May 6, 2003 gave recommendations for avoiding the problem and indicated that updated software to prevent the problem would be introduced.
Root Cause
Other
Distribution Pattern
Nationwide in the United States and worldwide.