STS-T Mobile x-ray system used with the STS-T lithotripter.
FDA Recall Z-0884-04 · Posted 2004-08-28
Advertisement
Recalling FirmMedstone International Inc
Recall StatusTerminated
Date Initiated2004-06-11
Date Terminated2004-09-01
Product Quantity30 unites
Reason for Recall
Units are defective in that the identification and certification labels were inadequate or missing. The beam limiting device was not appropriate for general purpose radiographic procedures and the beam limiting device could ot collimate down to 125 square centimeters or less as required.
Firm Action
Field Service Engineers will install a dual fixed-size colimator and console label and a new user manual will be provided.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.