ACCU-CHEK Tender I 13/110 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.

FDA Recall Z-0885-2015 · Posted 2014-12-30

Recall StatusTerminated
Product CodeFPA
Date Initiated2014-11-20
Date Terminated2018-01-10
Product QuantityTotal of 13,041 boxes

Reason for Recall

Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify

Firm Action

Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.