Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis

FDA Recall Z-0887-2016 · Posted 2016-02-25

Recall StatusTerminated
Product CodeKPO
Date Initiated2016-01-18
Date Terminated2017-08-14
Product Quantity848 (212 cases)

Reason for Recall

On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.

Firm Action

The firm, Fresenius Medical Care, issued an "URGENT-MEDICAL DEVICE RECALL" letter dated 1/14/2015 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your stock for the product; if found, discontinue use immediately; place all units in a secure, segregated area; contact your FMCNA Customer Service Team at 1-800-323-5188 for instructions how to return the affected lot (lot# 15STAC072), and promptly complete and return the attached reply form by Fax to: 781-699-9796 Attn: Quality Department or scan and email to: [email protected]. Fresenius will pick up and replace your product. If you have any additional medical concerns or questions, please contact Medical Information and Communication at 855-816-2309 or Website: www.fresenius-medinfo.com. Further information or support concerning this issue, contact the Customer Service Care Team at 800-323-5188 and reference the and Support Field Action Number.

Root Cause

Error in labeling

Distribution Pattern

US Distribution to states of: KS, LA, MS, OK, and TX.