Brius Patient Specific Brackets (components in a set of custom metal orthodontic devices); BRIUS , RX Only, Non-Sterile, Custom Made Device, ID: 52345400UL
FDA Recall Z-0889-2022 · Posted None
Recalling FirmBrius Technologies Inc.
Recall StatusTerminated
Product CodeDYJ
Date Initiated2022-02-07
Date Terminated2024-10-18
Product Quantity4
Reason for Recall
Pontics, brackets, and customized orthodontic appliances were sold and distributed prior to completing the standard design control process, being documented, and released.
Firm Action
On 2/28/22, the firm, Brius Technologies Inc., sent "URGENT: MEDICAL DEVICE RECALL" Letters (dated 2/21/2022) with an attached Acknowledgement and Receipt Response Form (Pontics, Patient Specific Brackets , and Customized Appliances), were sent to the respective Physicians via Fed-Ex first class trackable mail; a signed acknowledgement and receipt response was requested. Actions to be taken by the customer/user include the following: At this time, it is not recommended to remove the devices and interrupt the patient's treatment. However, physicians need to monitor their patients closely for adverse events.
Additional Actions for Physicians and Patients with Pontics include in the letter:
It is recommended that patients with Pontics be monitored for abnormal inflammation of tissues close to the location of the Pontics in the mouth. If the inflammation is thought to be an allergic response, it is recommended the removal of the device. Another cause of inflammation could be poor oral hygiene. In this case, it is not recommended to remove the Pontic.
Additional Actions for Physicians and Patients with Patient Specific Brackets:
All Brackets distributed prior to market release (7/22/21) have been removed from patients. No further actions are needed at this time.
Additional Actions for Physicians and Patients with Customized Appliances:
It is recommended that if the TAD (Temporary Anchorage Device) used with the customized device fails during treatment, the physician should remove the customized appliance and request a replacement appliance (that will not be customized to fit TADs).
If customers have any questions they should contact: Elisabeth Miller
[email protected],586-242-7718
Acknowledgment response forms should be completed and returned as follows:
PLEASE EMAIL COMPLETED RESPONSE FORM TO: [email protected]
OR MAIL TO:
Attn: Michael Gregory
2611 Westgrove Dr. Ste109
Carrollton, TX 75006
Root Cause
Device Design
Distribution Pattern
US Nationwide: CA, DC, MA, NJ, NY, TX
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.