Terminated

Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for Sale. Pre-printed tube: MS31071, Shelf carton: PL51091-08 Atrapro Antipruritic HydroGel is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including radiation dermatitis and atopic dermatitis.

FDA Recall Z-0890-2013 · Posted 2013-02-26

Recall StatusTerminated
Product CodeFRO
Date Initiated2013-02-12
Date Terminated2013-11-25
Product Quantity56,566 1.5 oz tubes

Reason for Recall

Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.

Firm Action

Oculus sent an Urgent Medical Device Recall letter dated February 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue distribution and segregate the affected product. Immediately remove the affected lots of Recall product from the inventory of sales representative, sub contracted distribution and or fulfillment centers. Segregate the affected product in a secure location for return or destruction. If the product was further distributed customers were instructed to notify healthcare professionals of this recall. Customers were also instructed to completed and return the Business Reply Card and fax to 1-415-462-5163. Once the firm receives the affected product the customer would be contacted by an Oculus Representative and provided a Returns Goods Authorization (RGA) number. For questions customers were instructed to call 1-707-559-2445. For questions regarding this recall 707-559-7191, ext 14.

Root Cause

Incorrect or no expiration date

Distribution Pattern

Nationwide Distribution - One distributor in PA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.