Recall StatusTerminated
Date Initiated2005-09-30
Date Terminated2007-01-16
Product Quantity40
Reason for Recall
Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.
Firm Action
Guidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patient management recommendations have not changed.
Distribution Pattern
Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.