Life Pulse High Frequency Ventilator (HFV)
FDA Recall Z-0893-06 · Posted 2006-05-23
Advertisement
Recalling FirmBunnell, Inc.
Recall StatusTerminated
Product CodeLSZ
Date Initiated2005-04-04
Date Terminated2007-11-27
Product Quantity297 units
Reason for Recall
Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing the solenoid in the valve to function sporadically.
Firm Action
Notified consignees by letter on 04/04/2005.
Root Cause
Other
Distribution Pattern
Nationwide, Military medical center, Canada, and Czech Republic
Official Source