Peel-Away Introducer set, 7.0 FR, Catheter introducer, includes sheath: 7.0 FR., 15.5 CM long, Wire: .035" DIA., 50 CM long, 3 MM J needles: 18 ga., 7.0 cm long. Accepts devices up to 7.0 French, sterile, Catalog Number: C-PLI-7.0-35-X-1. Cook Inc., Bloomington, IN 47404
FDA Recall Z-0893-2008 · Posted 2008-04-16
Advertisement
Recalling FirmCook, Inc.
Recall StatusTerminated
Product CodeDYB
Date Initiated2007-12-07
Date Terminated2008-04-15
Product Quantity1,881 of total products
Reason for Recall
Packaging Problem: The peel-away sheath does not peel uniformly or completely.
Firm Action
Cook Inc. initially notified consignees via FEDEX by an Urgent: Product Recall on 12/7/07. The recall was then expanded and consignees were notified on 12/18/08 by another Urgent: Product Recall. The letters instructed any distributors to inform their customers of recall, and all users to return the recalled products and enclosed Recall Response Form to the firm. For additional information, contact 1-800-346-2686.
Root Cause
Package design/selection
Distribution Pattern
Worldwide Distribution- including USA and countries of Argentina, Australia, Canada, Chile, Denmark, Japan, New Zealand and Singapore.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.