Symbia Intevo Bold and Symbia Intevo 6 SPECT/CT Systems
FDA Recall Z-0893-2022 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc.
Recall StatusTerminated
Product CodeKPS
Date Initiated2022-03-11
Date Terminated2023-04-14
Product Quantity7 devices
Reason for Recall
There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue.
Firm Action
Customer Advisory Notice Letter will be provided to each impacted consignee. Depending on the customer s capabilities, Siemens provides the letters via mail, e-mail, or hand delivery by a service engineer. The receipt of the letter will be documented, and additional attempts will be made until such receipt is acknowledged.
The safety advisory letter informs customers: You may continue to use your system ensuring to follow all instructions in the Operator Manual.
If any changes in system vibrations or noises are detected, then activate the Emergency Stop (E-Stop), discontinue use of the system immediately and contact your service engineer.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distribution OUS only to El Salvador, France, Germany, Japan, Taiwan, and UK.