C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293

FDA Recall Z-0896-2022 · Posted None

Recall StatusTerminated
Product CodeJKD
Date Initiated2021-12-21
Date Terminated2024-02-12
Product Quantity3 Kits

Reason for Recall

ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible

Firm Action

DRG issued letter via email on 12/28/21. Letter states reason for recall, health risk and Action to be taken by distributors and end-users: Please check your inventory for the respective lot, stop using these kits immediately and promptly quarantine these products. Only if you faced above mentioned issue of SO OD <1.2, please ask for replacement of the kits that were affected and provide your data. All unused kits of the affected lot will be replaced.Please reply immediately upon checking your stock/use to this notice to confirm. All replies should be received at once OR by January 7th, 2022. We sincerely apologize for the inconvenience this may have caused you and/or your Customer. Reply to email: [email protected] and copy [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

NY