GE Healthcare Innova 4100 and Innova 4100 with Bolus Chasing offered as an option. Digital Fluoroscopic Imaging System.

FDA Recall Z-0898-06 · Posted 2006-05-24

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeMQB
Date Initiated2006-03-06
Date Terminated2008-02-24
Product Quantity291 (213 domestic and 78 OUS)

Reason for Recall

Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient.

Firm Action

GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.

Root Cause

Other

Distribution Pattern

Nationwide including Puerto Rico and OUS to include: Australia, Belgium, Brazil, Bulgaria, Byelarus, Canada, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Honduras, Hungary, Indonesia, Iraq, Ireland, Israel, India, Iran, Italy, Jordan, Japan, Republic of Korea, Lebanon, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic of China, Thailand, Tunisia, Turkey, United Kingdom, Venezuela