REF CMS-CT6T225, 6F Triple Lumen CT PICC with M.S.T. Components Product Usage: Power injectable infusion PICC insertion kits

FDA Recall Z-0898-2013 · Posted 2013-02-28

Recall StatusTerminated
Product CodeLJS
Date Initiated2012-12-13
Date Terminated2013-07-31
Product Quantity785

Reason for Recall

The affected product was packaged with the incorrect introducer needle. The kit label indicates a "safety" needle is included; the kits were packaged with a non-safety needle.

Firm Action

Medcomp sent a notification letter dated December 13, 2012 via email to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions or concerns call 215-256-4201

Root Cause

Nonconforming Material/Component

Distribution Pattern

USA Nationwide Distribution including the state of PA.