Roche/Hitachi Glucose Hexokinase Reagent for use on the Hitachi Modular System; Catalog # 1929542 (11929542216). In vitro diagnostic.
FDA Recall Z-0899-06 · Posted 2006-05-25
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeCFR
Date Initiated2006-04-12
Date Terminated2006-10-16
Product Quantity269 kits
Reason for Recall
A contaminant will change the pH of the R2 bottle, causing deterioration of the enzyme, and resulting in incorrect or no results being reported.
Firm Action
A recall notification letter dated 4/12/06 was sent to each customer instructing them to discontinue use of this lot of product.
Root Cause
Other
Distribution Pattern
Nationwide.