Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.

FDA Recall Z-0899-2015 · Posted 2014-12-30

Recall StatusCompleted
Product CodeFMI
Date Initiated2014-11-21
Product Quantity10,000

Reason for Recall

Needle is not captured in the needle safety sheath in specific lots.

Firm Action

Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.

Root Cause

Process control

Distribution Pattern

US Nationwide Distribution in the states of AZ CA FL GA IL IN LA MA MN MO MS MT NC ND NJ NV NY OH OR PA RI TN TX VA WA WI WV