Terminated

Equinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kit

FDA Recall Z-0900-2017 · Posted None

Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeKWT
Date Initiated2016-10-25
Date Terminated2019-07-26
Product Quantity2,278 units

Reason for Recall

Disengaged screw head could potentially not be retained in the Torque Defining Screwdriver.

Firm Action

Exactech sent an Advisory Notice on October 25, 2016, to Exactech field agents in the U.S. Agents outside of the U.S. were notified on November 2, 2016. Field agents were notified of the recall in the Advisory Notice and asked to notify their accounts that may have the recalled product in stock. Agents were required to complete and return the Field Advisory Notice Response Form within 5 working days. For further questions, please call (352) 377-1140.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MD, ME, MN, MO, NC, NJ, NV, NY, OH, OK, OR, PA, SC, TN, VA, WA and WI., and to the countries of : Austria, Australia, Canada, Germany, France, Spain, India, Switzerland, Italy, South Korea, United Kingdom, Brazil, Luxembourg, Greece, Guatemala, Netherlands, South Africa, and Sweden