Terminated

BIO CAL(R) Temperature Controller, Models: (a) 370, (b) 370I Product Usage: The BIO CAL(R) (unit) is a thermal transfer module designed to control patient blood temperature during cardiopulmonary bypass procedures when heating or cooling of the blood is desired.

FDA Recall Z-0900-2018 · Posted None

Recall StatusTerminated
Product CodeDWJ
Date Initiated2018-02-08
Date Terminated2020-08-06
Product Quantity1980 units

Reason for Recall

Medtronic has been unable to develop a cleaning protocol to satisfy current industry concerns and expectations. As a result, an updated cleaning protocol will not be developed by Medtronic and it has been determined that the best course of action is to request BIO CAL users to discontinue use and dispose of BIO CAL devices.

Firm Action

Medtronic sent an Urgent Medical Device Recall dated February 2018 via UPS 2-Day.to their customers. The letter identified the affected product, problem and actions to be taken. Medtronic recommended that the affected device no longer be used for clinical procedures. Customers are asked to dispose of any devices in their possession per their normal equipment obsolescence procedures. Customer are also instructed to not resell any devices for clinical use. For questions contact your Medtronic Sales Representative.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution