Century Console Base, Model #043625051b. Intended for use as a platform for other ancillary equipment and disposable products in cardiopulmonary surgical procedures.
FDA Recall Z-0902-2013 · Posted 2013-03-01
Advertisement
Recalling FirmSalyer Prn Biomedical
Recall StatusTerminated
Product CodeDTQ
Date Initiated2013-01-31
Date Terminated2013-04-15
Product Quantity9 units
Reason for Recall
Concerns that the addition of the UPS Battery Backup to the Century Console Base Shelf did not comply with the FDA premarket notification regulations.
Firm Action
An Urgent: Medical Device Recall letter, dated 1/31/13, was sent to all the customers that received the Century Console Base Shelf option with the Linear/1.5k 115v UPS Battery Backup for the Century Heart Lung Machine. The letter informed the customers that Salyer PRN Biomedical will have a Salyer Biomedical technician return the customers' Century Console Bases to their original configuration without the Shelf option and restore the Linear/1.5k UPS Battery Backup to its original condition as a stand-alone unit with its own wheel base for ease of use. A refund or credit will be issued for the return of the product. Customers are instructed to contact Jeff Poland at (314) 497-5232.
Root Cause
No Marketing Application
Distribution Pattern
Distributed in the states of MA,CA, and WI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.